Step 1

              Starting the configuration process.               
    1. Turn on the autosampler. Screen will read “…STANDBY…”
    1. Press “ENTER/PROGRAM” button. Use arrows to highlight “CONFIGURE”. Press “ENTER/PROGRAM” button
  • This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the “ENTER/PROGRAM” button.
  1. Turn on the autosampler. Screen will read “…STANDBY…”

  2. Press “ENTER/PROGRAM” button. Use arrows to highlight “CONFIGURE”. Press “ENTER/PROGRAM” button

This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the “ENTER/PROGRAM” button.

1024

Step 2

              Configuring "set clock".               
  • The first option for configuring is “SET CLOCK”
    1. Press “ENTER/PROGRAM” to “SET CLOCK”. Use the numeric keys to set the time and date. You can use the arrows to move. When finished. Press “ENTER/PROGRAM” button.

The first option for configuring is “SET CLOCK”

  1. Press “ENTER/PROGRAM” to “SET CLOCK”. Use the numeric keys to set the time and date. You can use the arrows to move. When finished. Press “ENTER/PROGRAM” button.

Step 3

              Configuring "bottles and sizes".               
    1. Press “ENTER/PROGRAM” to configure “BOTTLES AND SIZES”.
  • Use arrows to select “PORTABLE” sampler. Press “ENTER/PROGRAM” to configure bottles and sizes.
  • Use arrows to select “24” botles. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “1000” mL for the “BOTTLE VOLUME”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “BOTTLES AND SIZES”.

Use arrows to select “PORTABLE” sampler. Press “ENTER/PROGRAM” to configure bottles and sizes.

Use arrows to select “24” botles. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “1000” mL for the “BOTTLE VOLUME”. Press “ENTER/PROGRAM” button.

Step 4

              Configuring "suction line".               
    1. Press “ENTER/PROGRAM” to configure “SUCTION LINE”.
  • Use arrow to select “3/8” inch for “SUCTION LINE I.D.” Press “ENTER/PROGRAM” button.
  • Use arrow to select “VINYL” for “SUCTION LINE” type. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “25” feet for “SUCTION LINE LENGTH”. Press “ENTER/PROGRAM” button.
  • The suction line length may change based on location and should be measured!
  1. Press “ENTER/PROGRAM” to configure “SUCTION LINE”.

Use arrow to select “3/8” inch for “SUCTION LINE I.D.” Press “ENTER/PROGRAM” button.

Use arrow to select “VINYL” for “SUCTION LINE” type. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “25” feet for “SUCTION LINE LENGTH”. Press “ENTER/PROGRAM” button.

The suction line length may change based on location and should be measured!

Step 5

              Configuring "liquid detector".               
    1. Press “ENTER/PROGRAM” button to configure “LIQUID DETECTOR”.
  • Use arrows to select “ENABLE” “LIQUID DECTOR”. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “3” “RINSE CYCLES.” Press “ENTER/PROGRAM” button.
  • Screen will read “ENTER HEAD MANUALLY?” Use arrows to select “NO”. Press “ENTER/PROGRAM” button.
  • Screen will read “RETRY UP TO X TIMES WHEN SAMPLING?” Use numeric keys to enter “3”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” button to configure “LIQUID DETECTOR”.

Use arrows to select “ENABLE” “LIQUID DECTOR”. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “3” “RINSE CYCLES.” Press “ENTER/PROGRAM” button.

Screen will read “ENTER HEAD MANUALLY?” Use arrows to select “NO”. Press “ENTER/PROGRAM” button.

Screen will read “RETRY UP TO X TIMES WHEN SAMPLING?” Use numeric keys to enter “3”. Press “ENTER/PROGRAM” button.

Step 6

              Configuring "programming mode".               
    1. Press “ENTER/PROGRAM” to configure “PROGRAMMING MODE”.
  • Use arrows to select “BASIC” mode. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “PROGRAMMING MODE”.

Use arrows to select “BASIC” mode. Press “ENTER/PROGRAM” button.

Step 7

              Configuring "calibrate sampler".               
    1. Press “ENTER/PROGRAM” to configure “CALIBRATE SAMPLER”.
  • Use arrows to select “DISABLE” “CALIBRATE SAMPLER”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “CALIBRATE SAMPLER”.

Use arrows to select “DISABLE” “CALIBRATE SAMPLER”. Press “ENTER/PROGRAM” button.

Step 8

              Configuring "start time delay".               
    1. Press “ENTER/PROGRAM” to configure “START TIME DELAY”. Press “ENTER/PROGRAM” button.
  • Screen will read “X MINUTE DELAY TO START.” Use numeric keys to enter “0” minutes. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “START TIME DELAY”. Press “ENTER/PROGRAM” button.

Screen will read “X MINUTE DELAY TO START.” Use numeric keys to enter “0” minutes. Press “ENTER/PROGRAM” button.

Step 9

              Configuring "enabling pin".               
    1. Press “ENTER/PROGRAM” to configure “ENABLE PIN”. Press “ENTER/PROGRAM” button.
  • Use arrow to select “NO” for “MASTER/SLAVE MODE?” Press “ENTER/PROGRAM” button.
  • Use arrow to select “NO” for “SAMPLE UPON DISABLE?” Press “ENTER/PROGRAM” button.
  • Use arrow to select “NO” for “SAMPLE UPON ENABLE?” Press “ENTER/PROGRAM” button.
  • Use arrow to select “NO” for “RESET SAMPLE INTERVAL?” Press “ENTER/PROGRAM” button.
  • Use arrow to select “NO” for “INHIBIT COUNTDOWN?” Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “ENABLE PIN”. Press “ENTER/PROGRAM” button.

Use arrow to select “NO” for “MASTER/SLAVE MODE?” Press “ENTER/PROGRAM” button.

Use arrow to select “NO” for “SAMPLE UPON DISABLE?” Press “ENTER/PROGRAM” button.

Use arrow to select “NO” for “SAMPLE UPON ENABLE?” Press “ENTER/PROGRAM” button.

Use arrow to select “NO” for “RESET SAMPLE INTERVAL?” Press “ENTER/PROGRAM” button.

Use arrow to select “NO” for “INHIBIT COUNTDOWN?” Press “ENTER/PROGRAM” button.

Step 10

              Configuring "event mark".               
    1. Press “ENTER/PROGRAM” to configure “EVENT MARK”. Press “ENTER/PROGRAM” button.
  • Use arrow to select “CONTINUOUS SIGNAL”. Press “ENTER/PROGRAM” button.
  • Use arrow to select “PUMP CYCLE”. Press “ENTER/PROGRAM” button.
  • It is critical this setting is configured in order for the sensor node to operate the autosampler!
  1. Press “ENTER/PROGRAM” to configure “EVENT MARK”. Press “ENTER/PROGRAM” button.

Use arrow to select “CONTINUOUS SIGNAL”. Press “ENTER/PROGRAM” button.

Use arrow to select “PUMP CYCLE”. Press “ENTER/PROGRAM” button.

It is critical this setting is configured in order for the sensor node to operate the autosampler!

Step 11

              Configuring "purge counts".               
    1. Press “ENTER/PROGRAM” to configure “PURGE COUNTS”. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “XX” for “PRE-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “XX” for “POST-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “PURGE COUNTS”. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “XX” for “PRE-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “XX” for “POST-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.

Step 12

              Configuring "tubing life".               
    1. Press “ENTER/PROGRAM” to configure “TUBING LIFE”. Press “ENTER/PROGRAM” button.
  • Screen will read “X PUMP COUNTS, WARNING AT XXX000”. Press “ENTER/PROGRAM” button.
  • Use arrows to select “NO” for “RESET PUMP COUNTER?”. Press “ENTER/PROGRAM” button.
  • Use numeric keys to enter “XXX000” for “PUMP COUNTS TO WARNING”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “TUBING LIFE”. Press “ENTER/PROGRAM” button.

Screen will read “X PUMP COUNTS, WARNING AT XXX000”. Press “ENTER/PROGRAM” button.

Use arrows to select “NO” for “RESET PUMP COUNTER?”. Press “ENTER/PROGRAM” button.

Use numeric keys to enter “XXX000” for “PUMP COUNTS TO WARNING”. Press “ENTER/PROGRAM” button.

Step 13

              Configuring "program lock".               
    1. Press “ENTER/PROGRAM” to configure “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
  • Use arrows to select “DISABLE” “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
  1. Press “ENTER/PROGRAM” to configure “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.

Use arrows to select “DISABLE” “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.

Step 14

              Configuring "sampler ID".               
    1. The next configuration option is “SAMPLER ID”. Skip this configuration step by using the arrows.
  1. The next configuration option is “SAMPLER ID”. Skip this configuration step by using the arrows.

Step 15

              Configuring "run diagonostics".               
    1. The next configuration option is “RUN DIAGNOSTICS”. Skip this configuration step by using the arrows.
  1. The next configuration option is “RUN DIAGNOSTICS”. Skip this configuration step by using the arrows.