Step 1
Starting the configuration process.
- Turn on the autosampler. Screen will read “…STANDBY…”
- Press “ENTER/PROGRAM” button. Use arrows to highlight “CONFIGURE”. Press “ENTER/PROGRAM” button
- This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the “ENTER/PROGRAM” button.
Turn on the autosampler. Screen will read “…STANDBY…”
Press “ENTER/PROGRAM” button. Use arrows to highlight “CONFIGURE”. Press “ENTER/PROGRAM” button
This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the “ENTER/PROGRAM” button.
1024
Step 2
Configuring "set clock".
- The first option for configuring is “SET CLOCK”
- Press “ENTER/PROGRAM” to “SET CLOCK”. Use the numeric keys to set the time and date. You can use the arrows to move. When finished. Press “ENTER/PROGRAM” button.
The first option for configuring is “SET CLOCK”
- Press “ENTER/PROGRAM” to “SET CLOCK”. Use the numeric keys to set the time and date. You can use the arrows to move. When finished. Press “ENTER/PROGRAM” button.
Step 3
Configuring "bottles and sizes".
- Press “ENTER/PROGRAM” to configure “BOTTLES AND SIZES”.
- Use arrows to select “PORTABLE” sampler. Press “ENTER/PROGRAM” to configure bottles and sizes.
- Use arrows to select “24” botles. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “1000” mL for the “BOTTLE VOLUME”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “BOTTLES AND SIZES”.
Use arrows to select “PORTABLE” sampler. Press “ENTER/PROGRAM” to configure bottles and sizes.
Use arrows to select “24” botles. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “1000” mL for the “BOTTLE VOLUME”. Press “ENTER/PROGRAM” button.
Step 4
Configuring "suction line".
- Press “ENTER/PROGRAM” to configure “SUCTION LINE”.
- Use arrow to select “3/8” inch for “SUCTION LINE I.D.” Press “ENTER/PROGRAM” button.
- Use arrow to select “VINYL” for “SUCTION LINE” type. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “25” feet for “SUCTION LINE LENGTH”. Press “ENTER/PROGRAM” button.
- The suction line length may change based on location and should be measured!
- Press “ENTER/PROGRAM” to configure “SUCTION LINE”.
Use arrow to select “3/8” inch for “SUCTION LINE I.D.” Press “ENTER/PROGRAM” button.
Use arrow to select “VINYL” for “SUCTION LINE” type. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “25” feet for “SUCTION LINE LENGTH”. Press “ENTER/PROGRAM” button.
The suction line length may change based on location and should be measured!
Step 5
Configuring "liquid detector".
- Press “ENTER/PROGRAM” button to configure “LIQUID DETECTOR”.
- Use arrows to select “ENABLE” “LIQUID DECTOR”. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “3” “RINSE CYCLES.” Press “ENTER/PROGRAM” button.
- Screen will read “ENTER HEAD MANUALLY?” Use arrows to select “NO”. Press “ENTER/PROGRAM” button.
- Screen will read “RETRY UP TO X TIMES WHEN SAMPLING?” Use numeric keys to enter “3”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” button to configure “LIQUID DETECTOR”.
Use arrows to select “ENABLE” “LIQUID DECTOR”. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “3” “RINSE CYCLES.” Press “ENTER/PROGRAM” button.
Screen will read “ENTER HEAD MANUALLY?” Use arrows to select “NO”. Press “ENTER/PROGRAM” button.
Screen will read “RETRY UP TO X TIMES WHEN SAMPLING?” Use numeric keys to enter “3”. Press “ENTER/PROGRAM” button.
Step 6
Configuring "programming mode".
- Press “ENTER/PROGRAM” to configure “PROGRAMMING MODE”.
- Use arrows to select “BASIC” mode. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “PROGRAMMING MODE”.
Use arrows to select “BASIC” mode. Press “ENTER/PROGRAM” button.
Step 7
Configuring "calibrate sampler".
- Press “ENTER/PROGRAM” to configure “CALIBRATE SAMPLER”.
- Use arrows to select “DISABLE” “CALIBRATE SAMPLER”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “CALIBRATE SAMPLER”.
Use arrows to select “DISABLE” “CALIBRATE SAMPLER”. Press “ENTER/PROGRAM” button.
Step 8
Configuring "start time delay".
- Press “ENTER/PROGRAM” to configure “START TIME DELAY”. Press “ENTER/PROGRAM” button.
- Screen will read “X MINUTE DELAY TO START.” Use numeric keys to enter “0” minutes. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “START TIME DELAY”. Press “ENTER/PROGRAM” button.
Screen will read “X MINUTE DELAY TO START.” Use numeric keys to enter “0” minutes. Press “ENTER/PROGRAM” button.
Step 9
Configuring "enabling pin".
- Press “ENTER/PROGRAM” to configure “ENABLE PIN”. Press “ENTER/PROGRAM” button.
- Use arrow to select “NO” for “MASTER/SLAVE MODE?” Press “ENTER/PROGRAM” button.
- Use arrow to select “NO” for “SAMPLE UPON DISABLE?” Press “ENTER/PROGRAM” button.
- Use arrow to select “NO” for “SAMPLE UPON ENABLE?” Press “ENTER/PROGRAM” button.
- Use arrow to select “NO” for “RESET SAMPLE INTERVAL?” Press “ENTER/PROGRAM” button.
- Use arrow to select “NO” for “INHIBIT COUNTDOWN?” Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “ENABLE PIN”. Press “ENTER/PROGRAM” button.
Use arrow to select “NO” for “MASTER/SLAVE MODE?” Press “ENTER/PROGRAM” button.
Use arrow to select “NO” for “SAMPLE UPON DISABLE?” Press “ENTER/PROGRAM” button.
Use arrow to select “NO” for “SAMPLE UPON ENABLE?” Press “ENTER/PROGRAM” button.
Use arrow to select “NO” for “RESET SAMPLE INTERVAL?” Press “ENTER/PROGRAM” button.
Use arrow to select “NO” for “INHIBIT COUNTDOWN?” Press “ENTER/PROGRAM” button.
Step 10
Configuring "event mark".
- Press “ENTER/PROGRAM” to configure “EVENT MARK”. Press “ENTER/PROGRAM” button.
- Use arrow to select “CONTINUOUS SIGNAL”. Press “ENTER/PROGRAM” button.
- Use arrow to select “PUMP CYCLE”. Press “ENTER/PROGRAM” button.
- It is critical this setting is configured in order for the sensor node to operate the autosampler!
- Press “ENTER/PROGRAM” to configure “EVENT MARK”. Press “ENTER/PROGRAM” button.
Use arrow to select “CONTINUOUS SIGNAL”. Press “ENTER/PROGRAM” button.
Use arrow to select “PUMP CYCLE”. Press “ENTER/PROGRAM” button.
It is critical this setting is configured in order for the sensor node to operate the autosampler!
Step 11
Configuring "purge counts".
- Press “ENTER/PROGRAM” to configure “PURGE COUNTS”. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “XX” for “PRE-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “XX” for “POST-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “PURGE COUNTS”. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “XX” for “PRE-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “XX” for “POST-SAMPLE COUNTS”. Press “ENTER/PROGRAM” button.
Step 12
Configuring "tubing life".
- Press “ENTER/PROGRAM” to configure “TUBING LIFE”. Press “ENTER/PROGRAM” button.
- Screen will read “X PUMP COUNTS, WARNING AT XXX000”. Press “ENTER/PROGRAM” button.
- Use arrows to select “NO” for “RESET PUMP COUNTER?”. Press “ENTER/PROGRAM” button.
- Use numeric keys to enter “XXX000” for “PUMP COUNTS TO WARNING”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “TUBING LIFE”. Press “ENTER/PROGRAM” button.
Screen will read “X PUMP COUNTS, WARNING AT XXX000”. Press “ENTER/PROGRAM” button.
Use arrows to select “NO” for “RESET PUMP COUNTER?”. Press “ENTER/PROGRAM” button.
Use numeric keys to enter “XXX000” for “PUMP COUNTS TO WARNING”. Press “ENTER/PROGRAM” button.
Step 13
Configuring "program lock".
- Press “ENTER/PROGRAM” to configure “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
- Use arrows to select “DISABLE” “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
- Press “ENTER/PROGRAM” to configure “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
Use arrows to select “DISABLE” “PROGRAM LOCK”. Press “ENTER/PROGRAM” button.
Step 14
Configuring "sampler ID".
- The next configuration option is “SAMPLER ID”. Skip this configuration step by using the arrows.
- The next configuration option is “SAMPLER ID”. Skip this configuration step by using the arrows.
Step 15
Configuring "run diagonostics".
- The next configuration option is “RUN DIAGNOSTICS”. Skip this configuration step by using the arrows.
- The next configuration option is “RUN DIAGNOSTICS”. Skip this configuration step by using the arrows.