Introduction
What you need
Step 1
Air and O2 Blender
- Remove the ribbon cable at the connection on the oxygen blender control board.
Remove the ribbon cable at the connection on the oxygen blender control board.
1024
Step 2
- Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
Step 3
- The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
To reassemble your device, follow these instructions in reverse order. Perform an OVP test to check proper installment.
NOTE: do not interchange control boards with non-socketed ones!