Introduction
What you need
Step 1
Air and O2 Blender
- Remove the ribbon cable at the connection on the oxygen blender control board.
Remove the ribbon cable at the connection on the oxygen blender control board.
1024
Step 2
- Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
Mark the plastic hoses attached to the oxygen blender (so you can put them back on the same way) and then remove.
Step 3
- The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
The air/oxygen blender has two screws close to the control board. Remove them and remove the unit.
Step 4
Air and O2 Inlet Manifold
- Remove the pneumatic and electrical connections labeled.
Remove the pneumatic and electrical connections labeled.
Step 5
- Remove the water trap bowls on the rear of the unit.
Remove the water trap bowls on the rear of the unit.
Step 6
- Remove the four screws and remove the manifold.
Remove the four screws and remove the manifold.
Step 7
- Remove the two screws holding the power supply to the manifold.
Remove the two screws holding the power supply to the manifold.
To reassemble your device, follow these instructions in reverse order. Calibrate and test the system to see if the replacement was successful
Cancel: I did not complete this guide.
Author
with 1 other contributor
William Burns
Member since: 04/06/2020
285 Reputation
2 Guides authored
Badges:
5
+2 more badges
Team
Cal Poly, Team S7-G12, Paton Fall 2020
Member of Cal Poly, Team S7-G12, Paton Fall 2020
CPSU-PATON-F20S7G12
2 Members
6 Guides authored